Back Back

Specifications

Toneohm 950


RESISTANCE

Ranges ohm, 200m ohm, 2 ohm, 2k ohm, 20k ohm
Accuracy ±4% in 200m ohm, 2 ohm, 2k ohm
±5% in 20k ohm
Ohm range High sensitivity uncalibrated, approximately 40m ohm full scale
Probe voltage 60mV maximum
Probe protection Momentary contact up to 250V
Indication Tone and meter on all ranges

TRACK CURRENT

Ranges 200mA, 2A, Trace
Accuracy 200mA, 2A ±15%
Trace Reading proportional to current when UNCALIBRATED illuminates.
Reading proportional to detected magnetic field strength.
Capable of detecting current flow with 200 ohm across DRIVE SOURCE.
Probe voltage 600mV maximum in 200mA, 2A
Probe protection Momentary contact up to 25V in 200mA, 2A
Indication Tone and meter on all ranges

TRACK VOLTAGE

Ranges 2mV, 20mV, 20V
Accuracy 120 ohm in 2mV, 20mV
1M ohm in 20V
Input resistance ±4% ±15µV
Probe protection Momentary contact up to 250V in 20V; 30V in 20mV, 2mV
Indication Tone and meter on all ranges

PLANE SHORTS

Indication Tone, uncalibated meter and LED fault direction arrows
Sensitivity Adjustable to compensate for differing plane resistance

DRIVE SOURCE

Output voltage 0 to 550mV, adjustable
AC in TRACE, DC in all other modes
Protected to ±30V

PLANE STIMULUS

Output voltage 550mV maximum

POWER REQUIREMENTS

200-250V or 100-130V @ 50/60Hz, 25VA

ACCESSORIES
(STANDARD)

Detachable needle probes
Detachable current trace/drive source probe
Detachable plane probe
Four detachable colour-keyed plane stimulus leads
Lightweight headphones
Operator manual
All probes stored under hinged cover

APPROVALS

Conforms to applicable European Directives and is CE marked
Polar Instruments Ltd. is certified to ISO 9001,
LRQA certificate number 932134.

NATO STOCK NUMBER

Contact Polar Instruments